在线播放亚洲第一字幕,免费看裸裸体美女啪啪无遮挡,亚婷婷洲AV久久蜜臀无码,精品人妻无码AV一区二区

技術(shù)文章Article 當(dāng)前位置: 首頁 > 儀器使用視頻 > 幾家不同公司的接觸角儀軟件采用停滴法并用Young-Laplace方程擬合測試接觸角的區(qū)別

幾家不同公司的接觸角儀軟件采用停滴法并用Young-Laplace方程擬合測試接觸角的區(qū)別

點擊次數(shù):2282    發(fā)布時間:2022-09-26 00:00:00

本視頻中演示了采用Young-Laplace方程擬合時各家接觸角儀公司軟件的區(qū)別。從測試結(jié)果來看,接觸角變化范圍從103-105度之間,屬于一個合理的變化范圍。這種變化由各種因素造成,如水的揮發(fā)等。

 

 

另可參考如下資料,我們發(fā)現(xiàn)國外的部分生產(chǎn)廠的接觸角儀軟件使用Young-Laplace擬合時,90-140度時,測值偏大5度,而大于140度時,測值偏大約10度。

We measured contact angle of same sample with 4 different software from Kruss, Dataphysics, Biolin and KINO. And we measured contact angle of sample by Biolin's sofware - OneAttension V2.4 at first because only this software cannot import images into software and calculate contact angle for imported image. Other sofwares such as DSA1.92 and DSA3.0 from Kruss, and SCA202 from Dataphysics and CAST3.0 from KINO can all import image into software and can calculate contact angle using different method such as circle fitting method, ellispe fitting method and Young-Laplace equation fitting method.

At this article, we just show you the calculated result by different company's software. Comparison of measured result maybe has some existed deviation due to our misoperation. You can contact us to obtain the recorded video for measuring process of contact angle.

And from the comparison of measured contact angle from different drop shape software, may safely draw a conclusion that:

1, When contact angle below 3°:  anti-dazzling screen must be used to get sharp drop shape and when using Young-Laplace equation fitting method, only sofware from Biolin (OneAttension V2.4) is failure to fitting the drop shape and get a acceptable contact angle.

2, When there exists some noise points or poor backgroud, tangent method based on parts of drop profile will  failure to fitting the drop shape and get a acceptable contact angle.

3, When contact angle between 90°-140°, calculated contact angle based on Young-Laplace equation fitting method from drop shape analysis software from Kruss, Dataphysics and Biolin will be larger than other mehod such as circle fitting and ellispe fitting about 3-5° when circle fitting or ellispe fitting can fitting the drop shape. And also this value based on Young-Laplace equation fitting method is larger than ADSA-RealDrop based method from KINO about 3-5°.

4, When contact angle larger than 140°,  calculated contact angle based on Young-Laplace equation fitting method from drop shape analysis software from Kruss, Dataphysics and Biolin will be larger than other mehod such as circle fitting and ellispe fitting about 8-10° when circle fitting or ellispe fitting can fitting the drop shape. And also this value based on Young-Laplace equation fitting method is larger than ADSA-RealDrop based method from KINO about 8-10°.

5, Calculated ontact angles by different method from KINO's sofware CAST3.0 are almost kept same when both of them fitting drop shape correctly.

 Please be noted that,

1, A basic principle is contact angle is calculated from tangent at contact point or curve radius at contact point. And when fitting drop profile with some equation of curve, if result of fitting is good, calculated contact angle will be almost same. In other words, Young-Laplace equation can be used to calculate contact angle when drop shape cannot fitting with circle equation due to effect of gravity. But if we used other equation of curve and fitting it correctly, the result will be kept almost same.

2, Red number means it is failure to fitting drop profile with current method. And "—" means current method cannot calculate any result automatically.  

Comparison of calculated contact angle based on different method and different company such as Kruss, Dataphysics, Biolin and KINO.

版權(quán)所有 © 2020 上海梭倫信息科技有限公司  ICP備案號: 滬ICP備05051428號-2